Unpacking The Device; Removing The Display And Control Unit; Positioning The Display And Control Unit - Sartorius Cubis MCA125P Series Operating Instructions Manual

Semimicro, analytical and precision balances
Hide thumbs Also See for Cubis MCA125P Series:
Table of Contents

Advertisement

5.3

Unpacking the Device

Procedure
Lift the device with the styrofoam padding out of the packaging .
t
Place the device in the styrofoam padding on its side .
t
Lift the styrofoam padding off the device .
t
NOTICE Glass breakage due to incorrect handling of the device! Only lift the
t
device by its base .
Place the device on its base .
t
Keep all parts of the original packaging, e .g . to return the device .
t
5.4

Removing the Display and Control Unit

5.4.1

Positioning the Display and Control Unit

The display and control unit can be removed . This enables the flexible setup of the
display and control unit at the workplace .
Tool:
1 Torx Allen key, T20
Material:
1 soft support base
Requirements
− The weighing pan and the associated components have not been set up .
− For a device with an analytical draft shield or flat glass draft shield: The side
panels and upper panel have not been fitted .
Procedure
Turn the device on its side and place it on the soft support base .
t
To loosen the control unit's retainer: Use the Torx Allen key to remove both
t
screws .
Remove the control unit and re-insert both screws into the threaded holes .
t
Installation
Cubis
®
MCA Operating Instructions
39