Profile / Weighing" Menu - Sartorius Cubis MCA125P Series Operating Instructions Manual

Semimicro, analytical and precision balances
Hide thumbs Also See for Cubis MCA125P Series:
Table of Contents

Advertisement

System Settings
7.5.4
Parameters in the "Weighing and Print Profile / Weighing" Menu
Parameter
Settings
Ambient conditions
Very stable
Stable
Unstable*
Very unstable
Application filter
Weighing (final readout)*
Dosing (filling mode)
Sensor mode
(low filtering)
Dynamic mode
(without filtering)
Stability
Very high accuracy
High accuracy
Medium accuracy*
Fast
Very fast
Very slow
* Factory setting for the "Standard Weighing" factory-set weighing profile
54
Cubis
®
MCA Operating Instructions
Explanation
Sets the ambient conditions to "very stable": Activates a fast change in the
weight values in the event of a load change with a high output rate .
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
Sets the ambient conditions to "stable" .
Recommended for the following work environment:
− Stable table
− Slight movement in the room
− Slight draft
Sets the ambient conditions to "unstable": Activates the delayed change in
weight values with a reduced output rate .
Recommended for the following work environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
Sets the ambient conditions to "very unstable": Activates a significantly
delayed change in the weight values and long wait for stability with a
further reduction in the output rate .
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
Activates a filter that enables a fast change in the display for very fast load
changes .
Display changes with minimal load changes (in the digit range) occur more
slowly .
Activates a filter that enables a very fast change in the display with
minimal load changes (e .g . when filling containers) .
Activates a weak but fast filter that always behaves in the same way for
load changes (e .g . when filling automated systems) .
Deactivates the active application filter .
Sets the stability to "very high accuracy" .
Sets the stability to "high accuracy" .
Sets the stability to "medium accuracy" .
Sets the stability to "fast" .
Sets the stability to "very fast" .
Sets the stability to "very slow" .