Preparing A Device With A Frame Draft Shield; Cleaning The Device - Sartorius Cubis MCA125P Series Operating Instructions Manual

Semimicro, analytical and precision balances
Hide thumbs Also See for Cubis MCA125P Series:
Table of Contents

Advertisement

Remove the weighing pan and all associated components from the weighing
t
compartment, e .g . shield plate, pan support .
9.2

Preparing a Device with a Frame Draft Shield

Procedure
Turn the device off .
t
1
Disconnect the device from the power supply . To do so, disconnect the power
t
supply cable from the wall outlet .
2
Remove the frame draft shield (1) and clean it with a brush or a damp cleaning
t
3
cloth .
Remove the weighing pan (2) .
t
4
Remove the shield plate (3) and pan support (4) .
t
9.3

Cleaning the Device

NOTICE
Corrosion or damage to the device due to unsuitable cleaning agents
Do not use corrosive, chloride-containing, or aggressive cleaning agents .
t
Do not use cleaning agents that contain abrasive ingredients, e .g . scouring
t
agents, steel wool .
Do not use solvent-based cleaning agents .
t
Check whether the cleaning agents used are compliant materials (see
t
Chapter "15 .4 Materials", page 88) .
Make sure that cleaning materials, such as cloths, are only slightly damp .
t
Procedure
NOTICE Malfunction or damage to the device due to the ingress of moisture or
t
dust
Remove dust and powdery sample residue with a brush or hand-held vacuum
t
cleaner .
NOTICE Corrosion or damage to the weighing pan due to unsuitable cleaning
t
agents
Wipe down the balance base, the shield plate, and the weighing pan with a
t
cleaning agent and a cloth . The cleaning agent must be suitable for stainless
steel and titanium .
Wipe the associated components of the weighing pan with a brush or a damp
t
cleaning cloth, e .g . shield plate, support pan .
Cleaning and Maintenance
Cubis
®
MCA Operating Instructions
77