Installation
46
Cubis
®
MCA Operating Instructions
Rotate the panel 180° so that the recess (1) in the panel points towards the
1
t
weighing module .
Feed the connection cable into the weighing chamber .
t
Insert the panel into the guide groove (1) .
t
1
Lift the locking tab (1) on the rear panel of the device and push down the panel .
t
Press the locking tab down and close it .
t
1

5.12 Acclimatization

When a cold device is brought into a warmer area: The temperature difference can
lead to condensation of humidity in the device (moisture formation) . Moisture in
the device can lead to malfunctions .
Allow the device to acclimatize for approx . 2 hours at the installation site .
t
Ensure that the device is disconnected from the power supply during that time .