If the isoCAL manual start function is set and the [isoCAL] button is displayed as
t
the predominant button in the operating display:
Press the [isoCAL] button .
t
If information is displayed stating that the weighing pan is loaded: Unload
t
the weighing pan .
The isoCAL function starts .
y
If the isoCAL function is complete: The device confirms the completion of the
y
calibration / adjustment process with an acoustic signal, and the calibration
report is displayed .
To close the calibration report and return to the previous display: Press the [OK]
t
button .

8.11 Internally Calibrating and Adjusting the Device

Requirements
The weighing pan is unloaded .
Procedure
Open the main menu .
t
Select the "Internal Adjustment" task .
t
The internal calibration / adjustment function is executed .
y
If automatic leveling is set for a model with motorized adjusting feet: The device
y
levels itself automatically .
If the calibration / adjustment function is complete: The device confirms the
y
completion of the calibration / adjustment process with an acoustic signal, and
the calibration report is displayed .
To close the calibration report and return to the main menu: Press the [OK]
t
button .

8.12 Weighing and Printing with ID Marking

8.12.1 Saving Values for Printouts

Every lot and every sample can be assigned an ID number . The ID numbers are saved
in the print memory and exported during the printing process .
Requirements
The prompt for the lot IDs and sample IDs is activated for the "Standard weighing"
task .
Procedure
Open the main menu .
t
Start the "Weighing" task: To do so, tap on the task .
t
The input field for the lot ID is displayed .
y
If a barcode scanner is connected to the device: Enter the lot ID in the entry field
t
or scan using the barcode scanner .
If no barcode scanner is connected to the device: Type the lot ID into the entry
t
field .
Press the [OK] button .
t
The weighing display is displayed .
y
Zero the device . In order to do so, press the [Zero] button .
t
Place the sample on the weighing pan .
t
Press the [Confirm] button .
t
The input field for the sample ID is displayed .
y
Operation