Function
Sequence of operations 1. Enter the menu with the supervisor code.
34
7.6. The number of averaged measurements
(08. menu item)
The instrument performs cca. 12 measurements in each
seconds. The displayed measurement result is generated
as the average of several measurements.
Here you can define the number of measurements that
should be used for calculating the averaged numerical
value. By increasing this number the display stability
increases, but the signal settling time becomes lower.
[Default factory setting: 16]
The way the code should be type in can be found in
Chapter 7.1. Typing the code in. You see on the display:

.
2. Keep stepping by pressing the ▲ button as long as you
see this menu item:
3. Enter the menu item by pressing the ENTER button.

4. You see this
blinking it signs that it can be modify.
5. Pressing the ▲ button select the number of
measurements to be averaged for the displayed numerical
value from the followings: 32, 64, 1, 2, 4, 8, 16. [factory
default: 16].
Note 1: When you choose 1 the displayed value equals to
tle last measured value.
Note 2: the settlig time after stabilizing the input current =
= number of averaging / 12 [seconds].
Number of averaging: 1
Number of averaging: 2
Number of averaging: 4
Number of averaging: 8
Number of averaging: 16
Number of averaging: 32
Number of averaging: 64

(Averaging = átlagolás).
on the display. The mnemonic is
Settling time: 0.1 second
Settling time: 0.2 second
Settling time: 0.4 second
Settling time: 0.7 second
Settling time: 1.4 second
Settling time: 2.7 second
Settling time: 5.4 second
DT4260 xx xx xx xx
20111006-V020111006