Datcon DT4260 Series Operating Instructions Manual page 55

Ac current meter
Table of Contents

Advertisement

DT4260 xx xx xx xx
Digital output test
Sequence of operations
Analog output test
Sequence of operations
20111006-V0
8. Keep stepping by pressing ▲ button as long as you see

this menu item:
output test).
9. Enter the menu item by pressing the ENTER button.

10. You see:
output (Limit), meaning the two underline: first and second
limit outputs are in off state.
11. Pressing the ▲ button, you can switch the outputs on or
off.:


: both are switched OFF.


: 1. ON; 2. OFF.


: 1. OFF; 2. ON.


: both are switched ON.
Rel1, Rel2 yellow indicators on the left side of the front
panel indicate the ON status of the limit outputs.
12. Press the MENU button, to exit from the menu item, and

you see this:
If you want finish test, continue from part Exit from menu
item.
13. Keep stepping by pressing ▲ button as long as you see

this menu item:
output test).
14. Enter the menu item by pressing the ENTER button.
15. You see this:
16. Pressing ▲ button you can increese the value of output
current in 1 mA steps in 0 - 20 mA range.
17. Press the MENU button, to exit from the menu item, and

you see this:
(Test: Digital Output = digital
on the display. Meaning L: limit
.
(Test: Analog Output = analog

(00 mA).
.