Function
Sequence of operations 1. Enter the menu by typing the supervisor code.
46
7.9. The number of averaged measurements
(07. menu item)
The instrument performs cca. 15 measurements in each
seconds. The displayed measurement result is generated
as the average of several measurements.
Here you can define the number of measurements that
should be used for calculating the averaged numerical
value. By increasing this number the display stability
increases, but the signal settling time becomes lower.
[Default factory setting: 8]
The way the code should be type in is described in Chapter
7.1. Typing the code in. You see on the display:
2. Keep stepping by pressing the ▲ button as long as you
see this menu item:
3. Enter the menu item by pressing the ENTER button.
4. You see this
5. Pressing the ▲ button select the number of
measurements to be averaged for the displayed numerical
value.
: The displayed value is the average of the last 8
measurements.
[Default factory setting]
(The display settling time after the stabilisation of the input
signal: 0.5 seconds.)

: The displayed value is the average of the last 16
measurements.
(The display settling time after the stabilisation of the input
signal: 1 second.)

: The displayed value is the average of the last 32
measurements.
(The display settling time after the stabilisation of the input
signal: 2 seconds.)

.