3.3. Settings

DT9500 can be adjusted through the 3 button front panel
keypad. All configuration parameters are stored in the
instrument EEPROM for unlimited period of time, even
when the power supply beeng switched off.
According to the default factory setting, the probe Pt100 is
connected in 4-wire mode. For the transmitter -200 ºC is
assigned to 4 mA, and +600 ºC is assigned to 20 mA. The
number of averaged samples is 8, the display refreshing
time is 0.5 sec
After the apparatus has been installed, it is functionable
immediately, without any further settings.
Re-scaling the transmitter, or changing other factory default
settings (e.g. setting the device to 3-wire operating mode)
could be performed simply and easily by using the press-
buttons on the front panel. The instrument doesn't need
any internal adjustment.

3.4. Storing and transporting

This instrument should be stored and transport in places
whose climatic conditions are in accordance with Chapter
10.1. Technical specifications, as described under the
title: Environmental conditions.
The packaging of DT9500 consist of enviroment-friendly,
recyclable cardboard is used to protect the instrument
against the impacts of normal stresses occurring during
transportation. The corrugated cardboard box is made from
environment-friendly, recyclable paper. The inner
protective material is polyfoam and nylon, which should be
disposed of via specialised recycling companies.