Datcon DT9500 Operating Instructions Manual page 85

Intrinsically safe temperature meter / transmitter
Table of Contents

Advertisement

DT9500
Alarm mode
20220103-V2
The limit output may work in Alarm mode too. This mode is
switched ON by the
default factory setting it is switched off (
Alarm mode is ON (
it has to switch on according to the measured and
displayed physical value, but it will not switch off even if the
value changes so that it should cause it to switch off. The
output can be switched off by pressing the ENTER button
only (acknowledgement).
Of course, if the condition that causes the output to switch
on persists, the limit output will switch on again when the
ENTER button is released.
(Similary to working of fire-alarm gear generates fire alarm.
Not suffice to definitive liquidation receipt signal alarming,
but must liquidate fire, generative cause.
After switching on the instrument
After switching on the instrument the limit outputs are
disabled as long as it performs at least three
measurements necessary for a complete averaging. This
takes a few seconds maximum. In this way can be avoid
false signalling due to transients that may appear in the
measuring loop.

parameter. According to the

), the limit output switches on when

). When the