Datcon DT9500 Operating Instructions Manual page 36

Intrinsically safe temperature meter / transmitter
Table of Contents

Advertisement

Hysteresis of the low
limit
Sequence of operations
View of the used terms
36
14. Pressing the ▲ button you can switch on or off the
minus sign.

= 123.4 ºC

= -123.4 ºC
15. After you have defined the value for the limit, press the
MENU button to exit from the sub-menu item, and you see

this:
.
16. Keep stepping by pressing the ▲ button as long as you
see this sub-menu item:
means: the hysteresis of the low limit value. Here you can
define a numerical value which provides the hysteresis of
the low limit value (LL) by having added to and
subtracted from it. Consequently, the actual hysteresis is
twice the value defined here.
Comment: according to this drawing, the limit output is in

the
operating mode.

. LH (Low Hysteresis)
DT9500
20220103-V2