Function
Sequence of operations 1. Enter the menu with the supervisor code.
20220103-V2
7.8. The temperature value belonging to the 20 mA
value (06 menu item)
The apparatus regulates the loop current between 4 and 20
mA in proportion with the measured temperature.
In this menu item any ºC values between -200,0 ºC and
600,0 ºC could be assigned to the 20 mA loop current.
[Factory default setting: 600,0 ºC]
Attention! In addition to selecting the temperature value
assigned to 20 mA, also the temperature value to be
assigned to 4 mA must be selected (The previous menu
point).
Read chapter 7.1. Writing the code in to see how the
code should be written in. On the display you see this:

.
2. Keep stepping with the ▲ button as long as you see this

menu item:
3. Enter the menu item by using the ENTER button.

4. The number
blinking. The blinking digit indicates which of the digits is
selected. The selection can be modified by using the ▲
button.
6. Using the ► button you can go to the next digit.
7. By using the ▲ button, you can increase the value of the
blinking digit:




,
,
,
,
8. If you have selected the last digit (the digit on the right
side), and you press the ► button, a blinking colon will
appear on the left side:
that the positive or negative sign is being selected, and you
can now change it by the ▲ button.
9. With the ▲ button you can switch the negative sign on
or off.

= 123.4 ºC

= -123.4 ºC
.
is visible on the display. Digit 4 is





,
,
,
,
,

. This colon indicates,
, etc.