Datcon DT9500 Operating Instructions Manual page 19

Intrinsically safe temperature meter / transmitter
Table of Contents

Advertisement

DT9500
Wiring plan, connecting
the limit outputs
(see also "Application
example")
Select connection cable
Select connection cable
for Ex applications
Inserting the cable into
the instrument
20220103-V2
Be careful the polarity of the cables.
Take note the suitability of the connecting cable.
We recommend the use of screened twisted pair cable.
The wire cross-section should be 0.25-1.5 mm
Take note of the corresponding installation regulations for
Ex applications. In particular, make sure that no potential
equalisation currents flow over the cable screen. In case of
grounding on both sides (for suppress the influence of high
frequency interference signals) this can be achived by use
of a capacitor (e.g. ceramic capacitor 1 nF, 1500 V) or
separate potential equalisation. The low frequency
potential equalisation currents are thus supressed, but the
protective effect against high frequency interference
signals remains.
1. Prepare the cable for the connection.
Remove approx. 30 mm of cable mantle, strip approx.
8 mm insulation.
2. Remove the instrument front cover (see chapter 4.3.).
3. Loosen compression nut of the cable entry.
4. Insert the cable into the instrument through the cable
entry.
2
.
19

Advertisement

Table of Contents
loading

Table of Contents