Specifications
Performance conditions
AWG5200 Series Technical Reference
This section contains the specifications for the AWG5200 series Arbitrary
Waveform Generators.
All specifications are typical unless noted as warranted. Warranted specifications
that are marked with the
To meet specifications, the following conditions must be met:
The instrument must have been calibrated/adjusted at an ambient temperature
between +20 °C and +30 °C.
The instrument must be operating within the environmental limits. (See
Table 35 on page 24.)
The instrument must be powered from a source that meets the specifications.
(See Table 33 on page 22.)
The instrument must have been operating continuously for at least 20 minutes
within the specified operating temperature range.
symbol are checked in this manual.