Brief procedures
32
6. From the Waveform List window, assign the waveform PV_Square.wfmx
to Channel 1.
7. Click the Setup -> Channel tab and enable Channel 1.
8. Press the AWG front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to enable the outputs (front panel light off).
9. Click the Play button on-screen or press the button on the front panel of the
AWG.
10. Check that the channel's waveform is properly displayed on the test
oscilloscope screen.
11. Press AWG the front panel All Outputs Off button (or click All Outputs Off
on the Home screen) to disable the outputs (front panel light on).
12. Repeat steps 3 through 11 until all channels are checked, modifying the
instructions with the channel number under test.
13. Disconnect the test setup.
AWG5200 Series Technical Reference