Emerson Daniel 3410 Series Maintenance And Troubleshooting Manual page 80

Gas ultrasonic flow meters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Meter repairs
August 2021
5. Remove the three flat head screws and split lock washers securing Acquisition
6. Remove module from connection board by lifting from each end simultaneously.
80
Figure 3-28: Acquisition Module cable and transducer wiring
A. Acquisition cable
B. Acquisition Module
C. Acquisition wiring terminal blocks
D. Acquisition Connection board
NOTICE
Ensure the transducer cables are labeled for the chord configuration.
Note
If Power is disconnected as instructed, the Barrier Cable and Transducer Wiring can
remain connected while removing Acquisition Module.
Module to Connection Board.
CAUTION
Do not remove Module at an angle as this may cause damage to connectors.
Maintenance and Troubleshooting manual
Maintenance and Troubleshooting manual
DAN-20057317