4.2.14 Batteries

1.
Check battery limits
Enter the BATTERY service menu:
Monitor Setup
Install Service (Password 16-4-34)
Service (Password 26-23-8)
Frame
Power Supply
WPM Battery
Smart Batt1
Smart Batt2
Check that all values are within reasonable limits.
NOTE: In order to update values for a desired battery, disconnect the other battery
shortly.
2.
Battery operation
Disconnect the power cord (without switching the monitor to standby).
Check that the monitoring continues normally. Check also the Command board LEDs:
- the green mains power LED is off
- the green battery LED is on
- the orange battery LED is off
3.
Battery charging
Reconnect the power cord.
Check that the monitoring continues without problems. Check also the Command board
LEDs:
- the green mains power LED is on
- the green battery LED is off
- the orange battery LED is on or off continuously (not blinking)

4.2.15 Network

1.
Connection to network
Check that the CAT-5 cable connector is clean and intact, then connect it to the monitor.
Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol
appears under the clock in the upper right hand corner of the screen. Also a message
regarding the connected Central should appear in the message field of the screen.
If the message regarding the Central does not appear, check the status of the network.
2.
Ethernet address
Enter the network service menu:
Monitor Setup
Install Service (Password 16-4-34)
Planned maintenance
3-29
Document no. 2044677-001