B30 Patient Monitor

4.1.2 Gases

Noise Meas
activates the noise
measurement.
Sample gain adj adjusts the sampling
pump gain, i.e. for
adjusting the sample flow
measurement.
Pump ctrl
A manual control for the
sampling pump.
Zero valve ctrl A manual control for the
zero valve.
Service Data
CO
% field
shows real-time
2
concentrations.
CO
noise-%
is standard deviation of
2
concentration.
CO
/channels A-E
2
mV field: signal is scaled
to mV.
Gain: User gain. It is
scaled as (User
gain)/(Factory gain).
O2, N2O, AA
Not in use
ID
Not in use
ID unrel.
Not in use
MAC
Not in use
Sample Flow
is calculated from differential pressure and adjusted by the module. Zero value as
measured during initialization when the pump is off. Gain: sample flow measurement can
be calibrated by adjusting the gain.
Ambient
Ambient pressure is measured continuously. Amb-Work: ambient pressure - sampling
system internal pressure.
Fall time
Not in use
CO
-O
Delay
Not in use
2
2
Pump
Can be toggled ON/OFF. PWM output 0-100% is shown. Pump voltage is also shown.
Lamp
The state, PWM control, and current of the lamp are shown.
Fan
Not in use
Zero Valve
Can be toggled between the measurement state (MEAS) and the zeroing state (ZERO).
Occl Valve
Not in use
Temp
Temperatures measured by the module from TPX, CPU, and OM.
Time after power on In minutes after power on.
5-14
.
Document no
2044677-001