Gas Sampling System Adjustment; Flow Rate Measurement - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
Figure 9
3.
For maximum accuracy, let the monitor warm up for 30 minutes. The menu item Gas
calibration remains gray as long as the message 'Calibrating gas sensor' is displayed.
4.
Press the
5.
Wait until the 'Zero ok' and then the 'Feed gas' messages appear after each gas on the
screen.
6.
Open the regulator and feed calibration gas until the message 'Adjust' appears, then
close the valve.
7.
Check that the displayed gas value matches the value on the calibration gas container.
NOTE: Adjust the O
is 20%).
NOTE: If an error occurs during calibration or if no gas is fed, the text 'Calibr. error' appears.
Push the ComWheel to perform a new calibration.
If adjustments are required:
Turn the ComWheel to highlight the first gas to be adjusted and then push the ComWheel.
Turn the ComWheel until the displayed value matches the desired value in the gas bottle and
push it again.
If the message 'Zero error' is displayed, press the
calibration procedure.
The time of the last calibration is shown at the bottom of the menu page.

5.1.2 Gas sampling system adjustment

For flow rate measurements, a flow meter with a low flow resistance and the capability to
measure low flow rates is required. A sampling line of normal length has to be connected to the
monitor as it has a considerable effect on the flow.

5.1.3 Flow rate measurement

If any flow rates are not correct, first replace the Mini D-Fend water trap, then recheck the
flows.
The sampling flow rate is measured by a flow meter at the sampling line. The flow rate should
be between 135 and 165 ml/min. The flow rate is adjusted in the Gases service menu with
Sample gain adj.
8-24
.
Document no
2044677-001
Connecting sampling line to the gas valve and feeding gas
Airway Gas
key and select Gas calibration.
percentage according to the calibration gas (for 755580 the right O2 value
2
Normal Screen
key and repeat the

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents