GE B30 Technical Reference Manual page 236

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
5.
Click Open License button, choose the license file you have. The CPU Serial Number and
Software Version will be automatically imported through the license file.
6.
Press Emergence button to start software download.
NOTE: When the emergence download for the main software is done successfully,
waiting for 3 to 5 minutes until the B30 monitor restart automatically.
Verification and configuration
1.
Perform Factory Reset.
Monitor Setup
After the Factory Reset is completed, the B30 monitor will restart automatically.
2.
Set the time and date.
Monitor Setup
3.
If necessary, set the monitor's network communication according to the used network
software.
Monitor Setup
Central station software S-CNET01 -> DRI Level = 2001
Central station software S-CNET02 -> DRI Level = 2002
Central station software L-NET03 or L-NET05 -> DRI Level = 2003
NOTE: If the DRI level is changed, the B30 monitor will restart automatically. The
change will take effect only after restart.
4.
Verify that the software has been installed correctly.
Monitor Setup
5.
Verify that the active language is installed and activated correctly
Monitor Setup
A-4
.
Document no
2044677-001
Install/Service (16-4-34)
Service (26-23-8)
Set/Test
Factory Reset
Time and Date
Install/Service (16-4-34)
Service (26-23-8)
Frame
Network
Dri Config
DRI level
Install/Service (16-4-34)
Service (26-23-8)
SW Version / Unit Id
Install/Service (16-4-34)
Service (26-23-8)
SW Settings
Country Settings