GE B30 Technical Reference Manual page 40

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
Net ID connector X8
9 pin female connector
5
9
Nurse Call (pin 8)
The nurse call signal is generated by the red and yellow alarms. When activated, the signal is
set to the high state and remains at the high state until the alarm situation is over or the
SILENCE ALARM key is pressed. The high state range is from 3.5 to 5 V, while the low state
range is from 0 to 0.5 V.
Serial port X9
9 pin male connector
1-26
.
Document no
2044677-001
Pin
1
2
3
1
4
6
5
6
7
8
9
Pin
1
2
3
4
5
6
7
8
9
Signal
NET_ID_CS TTL out
NET_ID_CLK TTL out
NET_ID_DO TTL out
NET_ID_DI TTL input
GND
+4.75 to +5.25V_out
GND (Reserved)
Nurse call, CMOS output
GND
Signal
GND
RXD, RS-232 input
TXD, RS-232 output
+4.75 to +5.25V
GND
N/C
RTS#, RS-232 output
CTS#, RS-232 input
N/C

Hide quick links:

Advertisement

Table of Contents