20. Quality of recording
Start/ Stop
Press the
acceptable.
21. Recording speed
RECORD WAVE
Press the
that the length of the recorded waveform scale is 18.7 cm (±1.5 cm).
Change the paper speed setting to 1 mm/s:
Record/Print
RECORD WAVE
Press the
of the scale is now 3.0 cm (±0.5 cm).
General
22. Electrical safety check
Perform an electrical safety check and a leakage current test.
23. Functioning after electrical safety check
Check that the module functions normally after performing the electrical safety check.
24. Final cleaning
Turn off the monitor, disconnect and clean the module.
Fill in all necessary documents.
key again and check that the quality of the recordings is
key again and this time wait until the recording stops. Check
- Record Waveforms - Paper Speed - 1 mm/s
key and wait until the recording stops. Check that the length
Service procedures
8-17
Document no. 2044677-001