GE B30 Technical Reference Manual page 132

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
Table 1
Value
Received packets (Statistics
In/Packets)
Transmitted packets (Statistics
Out/Packets)
Received bytes (Statistics
In/Bytes)
Transmitted bytes (Statistics
In/Bytes)
CRC errors (CRC)
Frame errors (Frame)
Transmission errors (Transm.)
BER errors (BER)
Internal errors (Intern.)
Missed packets (Missed)
FIFO errors (FIFO)
Overrun errors (Overrun)
5-6
.
Document no
2044677-001
Ethernet service data
Usage
Total number of received packets since
last cold start.
Total number of transmitted packets
since last cold start.
Total number of received bytes since
last cold start.
Total number of transmitted bytes since
last cold start.
Number of received packets with
incorrect checksum.
Number of received packets with
incorrect frame structure.
Number or errors in packet
Transmission.
Internal error of the network board.
Number of received packets lost due to
overload.
Internal error of the network board.
Practically the same as above.
Notes
Refers to physical layer
problems. An erroneous packet
has often both frame and CRC
errors.
Must always be 0.
Must always be 0.
Must always be 0.
Must always be 0.

Hide quick links: