GE B30 Technical Reference Manual page 41

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

Main power X3
Mains connector
Module connector X4
5 pin connector
Defib & IABP sync connector X5
Female mini din7 connector Pin
DIRECT_ECG: Output signal follows first user lead (ECG1 lead) including augmented leads. If
pace is on, there is an enhanced pacer pulse (5 volts at 2 milliseconds) when pacemaker
triggering is detected.
In the event of a LEAD FAIL, refer to the following:
Lead Failed
Right arm
Left arm
Left leg
Chest
On LEADS OFF situation ECG analog output is flat line (0V).
PRESSURE_OUT: The pressure labeled ART is sent to the analog out. If no pressure is labeled
ART, P1 is sent to the analog out. Output is set to flat line (0V) if the pressure channel is not
zeroed.
Pin
Signal
L
Live
PE
Protected earth
N
Neutral
Pin
Signal
1
GND
2
Vmod 13.8 - 16 V
3
Data +
4
Data -
5
Shield
Signal
1
Defib_sync_out
2
Reserved
3
Analog GND
5
Digital GND
6
7
Blood Pressure Analog out
8
ECG Analog out
ECG Analog Out
III
II
I
I
Frame functional description
Document no. 2044677-001
1-27