Pulse Oximetry; Temperature; Invasive Blood Pressure - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
ST filter
Heart rate from ECG
Range30 to 250 bpm
Accuracy±5 bpm or ±5%, whichever is greater
Resolution1 bpm
ST levels (in main software)
ST level range-9 to +9 mm (-0.9 to +0.9 mV)
Resolution0.1 mm (0.01 mV)
Pacemaker pulse detection
Detection level
Pulse duration

2.2.3 Pulse oximetry

Measurement range
Calibration range
Accuracy
Display resolution
The monitor is calibrated against functional oxygen saturation SpO
Pulse rate from Pleth
Measurement range
Accuracy
Resolution
Display averaging
Adjustable pulse beep volume.
Pleth waveform
Scales
Start up scale is 20 mod% if AUTO is not selected to be the default setting.

2.2.4 Temperature

Measurement range
Measurement accuracy
Display resolution
Temperature test
Probe type

2.2.5 Invasive blood pressure

Measurement range
Measurement accuracy
Zero adjustment range
Zero time
Transducer sensitivity
Digital display
Resolution
1 Accuracy is based on deep hypoxia studies with volunteered subjects during motion and non-motion conditions over a
wide range of arterial blood oxygen saturation as compared to arterial blood CO-Oximetry. Accuracy may depend on
the sensor used, please refer to the instructions for use in the accessory package.
7-4
.
Document no
2044677-001
1
0.05 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.05 to 40 Hz (-3 dB, with 60 Hz reject filter)
2 to 700 mV
0.5 to 2 ms
0 to 100%
70 to 100%
100 to 70%, ±2 digits
69 to 0%, unspecified
1 digit = 1% of SpO
2
30 to 250 bpm
±5 bpm or ±5%, whichever is greater
1 bpm
5-20 s
2, 5, 10, 20, 50 mod%, Auto
10 to 45 °C (50 to 113 °F)
±0.1 °C
0.1 °C at 25-45 °C with reusable probes
automatic (every 10 min.)
compatible with YSI 400 series
-40 to 320 mmHg
±5% or ±2 mmHg, whichever is greater
±150 mmHg
less than 15 s
5 µV/V/mmHg
±1 mmHg
func.
2