Introduction - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

1

Introduction

If a problem occurs during the functional examination, check the components of the monitor
according to the below troubleshooting table. If the problem persists, please refer to the
following detail troubleshootings.
Problem
Nothing functions
E-PSM(P)W module does
not function
N-F(C)(REC) module does
not function
What to do
Unplug and re-plug the Power Cord. Also confirm that the cable is intact.
Confirm that the fuses are intact.
Remove and replace the module.
Confirm that the desired parameters are configured to be displayed.
Confirm that 'Occlusion' or 'Calibrating Gas Sensor' messages are not displayed.
Confirm that a D-fend water trap and a sample tube are attached.
Confirm that the desired parameters are configured to be displayed.
Remove and replace the module.
Introduction
4-1
Document no. 2044677-001