B30 Patient Monitor
Temperature measurement
2.
Check the temperature channels with a patient simulator.
Check that temperature measurement information is shown on the monitor screen as
configured on the simulator.
Invasive blood pressure measurement
3.
Check the function of the module and side panel membrane keys.
4.
Check the InvBP channels with a patient simulator.
5.
Zero the InvBP channels and check that the values and waveforms correspond to the
simulator settings.
SpO2 measurement
6.
Connect an SpO2 finger probe to the module. Check that the message 'Probe off' is shown
when the probe is not connected to a finger.
7.
Attach the SpO2 probe to your finger. Check that a reading of 95-99 and a pleth
waveform appear on the screen
Non invasive blood pressure measurement
8.
Check the function of the module and side panel membrane keys.
9.
Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that
the module identifies the cuff, i.e. the text ëAdultí appears in the NIBP digit field for a short
time.
Check that the module gives a reasonable measured result.

3.2.7 Recorder

1.
Press the
selected waveforms. Press the
2.
Check that the quality of the recordings is acceptable.

3.2.8 Network connection

1.
Check that the CAT-5 cable connector is clean and intact, then connect it to the Network
connector on the backside of the monitor.
3-16
.
Document no
2044677-001
Start/Stop
sidepanel key and check that the module starts recording the
Start/Stop
sidepanel key again to stop recording.