B30 Patient Monitor
To remove the interconnection and
UI board
To remove the network connector
board unit
To remove the CPU board
6-16
.
Document no
2044677-001
-
Remove the three screws (T10) holding the plastic flex board
cover.
-
Remove the interconnection and UI board carefully from the
frame.
-
Remove the screw (T10) holding the unit to the frame.
-
Disconnect the cable coming from the Network connector
board
-
Remove the Network connector board
-
Disconnect all the rest cables from the CPU board.
-
Remove the eight screws (T10) holding the CPU board to the
frame.
-
Lift the CPU board carefully up and remove it from the
frame.