Functional Inspection; General; Display - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

3.
Checnk external parts:
4.
Reattach the module cover and check that the latch is moving properly.
5.
Module installation
Plug in the module. Check that it goes in smoothly and locks up properly
6.
Module recognition
Check that the module is recognized, i.e. all the needed parameter information, except
invasive blood pressure, starts to show on the screen.

4.2 Functional inspection

4.2.1 General

1.
Connect the mains power cord
Check that the Mains power LED is lit.
2.
Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs turn on shortly, normal
start-up sound is heard from the loudspeaker and the monitoring screen appears.
No error messages should appear on the screen.
3.
Configure the screen for the parameters that are connected.
Check that all the connected modules are recognized, i.e. the required parameter
information is shown on the screen.
If connected, the recorder should record two lines of start-up information.
Preset the measurement settings for those parameters that are connected, for example:
Print/Recorder
Invasive Pressures
Others
Others
4.
Check that the monitor goes to battery use if the main cord is disconnected.

4.2.2 Display

1.
Check that the picture on the screen is correct
2.
Check the display brightness
Select
the front cover and the front panel sticker are intact
all connectors are intact and attached properly
the module box and latch are intact
- Record Waveforms - Waveform 1 - ECG1
- P1 'ART' Setup - Label - ART
- P2 'CVP' Setup - Label - CVP
- SPO2 Setup - Pleth Scale - AUTO
-
Resp Setup
- Size - 1.0
- Resp Rate Source - AUTO
- Measurement - ON
- Detection Limit - AUTO
- Waveform 2 - P1
- Waveform 3 - P2
Planned maintenance
3-19
Document no. 2044677-001