4 Preparation and operation
Requirement
Result
4.7.12 Changing the ventilation mode in CPR mode
Requirement
88
EN
MEDUMAT Standard
If the trigger is still not functioning properly in CPR CCSV mode in
spite of reducing the trigger and increasing the PEEP, you can
switch from CPR CCSV mode to CPR IPPV mode
(see "4.7.12 Changing the ventilation mode in CPR mode",
page
88).
Interrupting ventilation in CPR CCSV mode
During the analysis of the defibrillator, you can pause ventilation in
order to avoid ventilation artifacts in the analysis.
The device is switched on
The CPR CCSV mode is set.
1. Pause chest compressions.
The ventilation is paused for as long as the chest compressions are
continued (up to the set back-up ventilation time at most).
This chapter describes how you can switch between CPR Manual,
CPR IPPV and CPR CCSV when in CPR mode.
The device is switched on.
A CPR mode is set.
1. Select CPR mode field using the right-hand navigation knob.
2