7 Description of the modes
7.3.3 CPAP mode
Pressure
pMax
CPAP / PEEP
148
EN
MEDUMAT Standard
Description
Abbreviation
Long form
Type
Requirement
Ventilation parameters
Left-hand navigation knob
Central navigation knob
Right-hand navigation knob
CPAP
1/Freq. (spontaneous)
The set value CPAP/PEEP is used to increase the pressure level of
respiration in order to raise the functional residual capacity (FRC)
of a spontaneously breathing patient. The patient is able to
breathe spontaneously without any restriction at the set pressure
level. The CPAP mode is used exclusively on patients with adequate
spontaneous respiration.
In principle, the pressure is set at the end of expiration (PEEP). The
set maximum ventilation pressure (pMax) ensures the safety of the
patient.
2
CPAP
Continuous Positive Airway Pressure
Pressure-controlled
None
-
-
•
PEEP
•
pMax
•
Emergency mode
CPAP
1/Freq. (spontaneous)
Time