4 Preparation and operation
Result
4.4.5 Connecting MEDUtrigger
62
EN
MEDUMAT Standard
5. With disposable hose systems: Detach the blanking plug on the
CO
connection and connect the CO
2
CO
connection.
2
When doing so, note: Tighten the screw connection by hand.
To minimize the dead space, you can also connect the CO
measuring hose to a breathing system filter with gas connection
(e.g., WM 22162).
6. With the reusable hose system: Guide the CO
with the rest of the measuring hose system and ventilation
hose into the hose protection sleeve of the patient hose
system.
7. If necessary: Activate the capnography option
(see "6.3.9 Options", page
The CO
measuring hose is connected to the patient hose system.
2
The operational readiness of MEDUtrigger is indicated by 2 green
LEDs on MEDUtrigger. If MEDUtrigger is connected to the device
and the green LEDs on MEDUtrigger are lit, you can trigger
mechanical breaths manually by pressing the MEDUtrigger button.
2
measuring hose to the
2
measuring hose
2
139).
2