9 Function check
Result
212
EN
MEDUMAT Standard
4. Briefly press the On/Off button
without an oxygen supply.
The remaining oxygen is flushed out of the device.
5. Press and hold the On/Off button
to switch off the device.
6. Replace leaky components.
7. Check the system for leaks once more
system for leaks", page
8. If necessary: Look for other leaks and replace leaky
components.
9. If the leak cannot be rectified, have the device repaired.
The leak in the system has been rectified.
2
and operate the device
for at least 2 seconds
(see "9.6 Checking the
211).