6 Operator menu
Parameter
etCO
/
(only
2
with capnography
option)
Apnea mode (only if the pressure-controlled ventilation modes option is activated)
Height
Vt per kg bodyweight
PEEP Warning
138
EN
MEDUMAT Standard
Possible values
IPPV
BiLevel + ASB
4 ml/kg - 10 ml/kg
1 mbar - 21 mbar
Setting height*
Depending on the height selected (tidal volume Vt in ml per kg
bodyweight) the height which can be set is restricted to the
following minimum values:
Tidal volume Vt in ml per kg
bodyweight
4
5
6
7
8
9
10
2
Description
Here you can determine whether an
alarm should be output in the event
of rising or dropping end-expiratory
CO
.
2
Here you can set the apnea
ventilation mode for the CPAP and
CPAP + ASB modes.
Here you can set the tidal volume in
milliliters per kilogram body weight.
In the process, a variable is used to
convert the height to a tidal volume
(see "14.2 Calculation of body
weight on the basis of body height",
page
258).
Here you can set a limit value for the
positive end-expiratory pressure. A
warning is then given on the display
if this value is reached or exceeded.
In this case, the PEEP field in the
bottom right of the display turns red.
Minimum height which can be
set in cm
90
80
70
65
60
55
50