Requirement
Result
210
EN
MEDUMAT Standard
Risk of injury due to inoperational device!
Operation of the device after a failed function check may result in
injury to the patient.
⇒ Only operate the device after it passes the function check.
The function check ended with Device is not ready for use.
Precise information on the individual tests in the function check
can be found in the file fcheck
checks", page
259).
1. Check the components named in the instructions on the
display and replace if necessary.
2. Repeat the function check.
3. If the function check ends with Device is not ready for use
again: Contact your authorized dealer or
WEINMANN Emergency.
The patient valve of the reusable hose system is dismounted
(see "8.4 Disassembly of the reusable hose system", page
1. Check all parts of the patient valve for external damage.
If necessary: Replace damaged parts.
2. Check the PEEP control diaphragm and inspect the check valve
diaphragm:
If the diaphragm is torn, wavy, distorted or sticky: Replace the
diaphragm.
3. Assemble reusable hose system
reusable hose system", page
The patient valve of the reusable hose system has been checked
and is ready for use.
2
(see "14.3.1 Recorded function
(see "8.11 Assembling
190).
177).