6 Operator menu
Parameter
Infant/Child/Adult
Emergency mode (only if
the pressure-controlled
ventilation modes
option is activated)
Vt
pInsp (only if the
pressure-controlled
ventilation modes
option is activated)
Freq.
I:E
PEEP
∆ pASB (only with
activated flow
measurement + ASB
option)
134
EN
MEDUMAT Standard
In the Presets patient submenu, you can determine which presets
are assigned to the ventilation parameters of the different patient
groups:
Risk of injury from different alarm presets in the same or
similar devices!
Different alarm presets in the same or similar devices in different
application areas can confuse the user and result in injury to the
patient.
⇒ Select the same alarm presets in the same or similar devices.
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
monitoring are output in the additional functions CPR
2
and RSI. This can injure the patient.
⇒ Notify the user if the alarms in the additional functions CPR and
RSI have been deactivated.
⇒ Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
Possible values
IPPV
BiLevel + ASB
50 ml - 2000 ml, in 50 ml increments
3 mbar - 60 mbar
5/min - 50/min
1:4 to 4:1
0 mbar - 30 mbar
0 mbar - 30 mbar
2
Description
Here you can choose between IPPV
mode or BiLevel + ASB as the
emergency ventilation mode per
patient group.
Here you can set the tidal volume.
Here you can set the inspiratory
pressure.
Here you can set the frequency.
You can specify the inspiration to
expiration ratio here.
Here you can set the positive end-
expiratory pressure.
Here you can set the pressure
support.