Disruption or failure of therapy due to unsuitable cleaning
agents and disinfectants!
The use of incorrect cleaning agents and disinfectants can cause
the device to malfunction. This can result in serious or life-
threatening injury to the patient.
⇒ Never clean the device and accessories with bleach, bleach
solutions or phenolic compounds.
⇒ Only use the cleaning agents and disinfectants recommended
in these instructions for use
disinfection plan", page
Loss of mechanical or electrical safety resulting from
reprocessing of the device and accessories with unsuitable
cleaning agents and disinfectants!
Using incorrect cleaning agents and disinfectants might cause
damage to the surface of the device and/or accessories, as well as
impairing electrical and insulating properties. This can result in
severe or life-threatening injury to the user and patient.
⇒ Only use the cleaning agents and disinfectants recommended
in these instructions for use
disinfection plan", page
Disruption or failure of therapy due to liquid in the hose
system after hygienic reprocessing!
Drips in the reusable measuring hose system or FlowCheck sensor
can distort the measurement results. This can result in serious or
life-threatening injury to the patient.
⇒ Following hygienic reprocessing of the hose system, allow all
hose system components to dry thoroughly.
8 Hygienic reprocessing
(see "8.12 Cleaning and
196).
(see "8.12 Cleaning and
196).