Weinmann MEDUMAT Standard2 Instructions For Use Manual page 187

Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

8.7.1 Immerse disinfecting the reusable measuring
hose system
Risk of infection and contamination due to insufficient
hygienic reprocessing of the measuring hose system!
Rinsing the measuring hose system in alternating directions does
not guarantee any bacteria reduction and may injure the patient.
⇒ Only rinse measuring hoses in one direction.
The principle described only applies to the following parts of the
reusable measuring hose system:
Requirement
1. The disinfectants, doses, and exposure time for the individual
2. Prepare the immersion disinfection solution in accordance with
3. Connect a sterile disposable syringe (20 ml) to one end of a
4. Draw the immersion disinfection solution up through the hose
5. Disconnect the disposable syringe from the hose.
6. Immerse the hose in the disinfection solution.
Pressure-measurement hose
PEEP control hose
Measuring hose system connector
The reusable measuring hose system is disconnected from the
reusable hose system
hose system", page
The reusable measuring hose system has been cleaned
manually.
parts can be found in the cleaning and disinfection plan
(see "8.12 Cleaning and disinfection plan", page
the specifications of the disinfectant agent manufacturer.
hose.
into the disposable syringe by means of suction until both are
completely full.
When doing so, note:
All surfaces and lumens must be completely wetted.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
8 Hygienic reprocessing
(see "8.4 Disassembly of the reusable
177).
MEDUMAT Standard
196).
2
EN
187