In this menu you can set the scale for the time axis of the etCO
trend. The following settings are possible: auto, 5 min., 10 min.,
30 min., 60 min.,120 min. At a time setting of 5 minutes or
10 minutes, the device records the determined value every
15 seconds. At a setting of 30, 60 or 120 minutes, an average
value is recorded every 30 seconds.
In the "auto" setting, the x axis is scaled automatically depending
on the duration of the application.
The unit of the etCO
selected in the operator menu under the menu item Device
Configuration
5.3.7 Bluetooth (only with Bluetooth data
transmission option)
5-8 Bluetooth submenu
If the Bluetooth data transmission option is enabled and
activated in the operator menu
page
100), you can enable and disable the Bluetooth
here. If the Bluetooth
documentation system can connect to the device to retrieve
operating data.
trend is based on the CO
2
(see "6.3.7 Device configuration", page
(see "4.14 Enabling options",
®
connection is activated, an operation
MEDUMAT Standard
5 User menu
2
unit. This can be
2
130).
®
connection
2
EN
117