Weinmann MEDUMAT Standard2 Instructions For Use Manual page 85

Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

3. To start continuous ventilation again: Press the
Result
Ventilation pauses.
When the countdown reaches zero, ventilation automatically
restarts.
4.7.11 Performing ventilation in CPR CCSV mode
Delay in treatment due to simultaneous metronome outputs
from the ventilator and the defibrillator!
If the ventilator is used together with a defibrillator
(MEDUCORE Standard or MEDUCORE Standard
emit a metronome sound, the simultaneous metronome outputs
from both devices may confuse the user and cause delays in
treatment.
⇒ Where the ventilator and defibrillator are used at the same
time, switch off the metronome sound of
MEDUMAT Standard
⇒ Do not perform resuscitation without a metronome.
4 Preparation and operation
2
.
MEDUMAT Standard
field twice.
2
), which can also
2
EN
85