4 Preparation and operation
Requirement
86
EN
MEDUMAT Standard
Treatment delays due to triggered alarms for spontaneously
breathing patients which do not apply in the this application
scenario!
In CPR CCSV mode, triggered etCO
and result in delays in treatment.
⇒ Switch off etCO
menu.
Treatment delay due to impaired detection of chest
compressions due to obstructed airways!
In CPR CCSV mode, chest compressions are detected as a result of
pressure changes in the airways. If the airways are obstructed, this
detection is impaired, which can lead to delays in treatment.
⇒ Check the setting of the PEEP and the trigger in CPR CCSV
mode and adjust if required.
The device is switched on.
The CCSV option is activated.
1. Briefly press the CPR button
Depending on the preset in the operator menu, the device
switches to CPR CCSV mode or CPR Manual mode (Operator
menu | Presets patient | CPR mode | Start mode).
2. If
is already set in the operator menu: Continue
ventilation via chest compressions in CPR CCSV mode:
or
If Manual is already set in the operator menu: Use the right-
hand navigation knob to switch to
(see "4.7.12 Changing the ventilation mode in CPR mode",
page
88).
2
alarms can confuse the user
2
alarms for CPR CCSV mode in the operator
2
.
CCSV