8 Hygienic reprocessing
Manual
cleaning
(only
Part
necessary
with visible
soiling)
Disposable hose system
Disposable hose
system
Disposable
measuring hose
system
Hose clips
FlowCheck
Disposable item, do not reuse. Dispose of correctly
sensor
(disposable)
Water filter
CO
measuring
2
hose (only with
capnography
option)
FlowCheck
Wipe down with
sensor
neodisher
connection line
MediClean forte
FlowCheck
(Dr. Weigert).
sensor
Dosage:
connection line
10 ml/l,
with
Duration: treat
MEDUtrigger
all surfaces at
MEDUtrigger
least twice until
with connection
visibly clean
line
200
EN
MEDUMAT Standard
Wipe
disinfection
®
Wipe down with
Incidin™
Oxywipe S
(Ecolab)
The applicable instructions are those in the instructions for use
from the manufacturers of the individual components or parts.
Observe these instructions for use.