12 Storage

12.4 Storing the battery

Requirement
Result
230
EN
MEDUMAT Standard
The device and the battery have been cleaned and disinfected
(see "8.3 Preparing hygienic reprocessing", page
The battery is fully charged.
If available: The replacement battery is fully charged.
1. Insert the battery in the battery compartment and store the
device in a dry place.
or
Store the battery in a dry place outside of the device.
Material damage due to prolonged storage of the battery
without recharging!
Storing the battery for a prolonged period of time without
recharging can result in the rapid shutdown of and irreparable
damage to the battery.
⇒ When the battery is stored in the device without a power
connection: Charge battery every 3 months.
⇒ If the battery is not stored in the device: Charge battery with
SN < 20000 every 5 months and battery with SN ≥ 20000
every 9 months.
2. Charge battery at regular intervals:
Type of storage
In device without power connection
Not in device
The battery is stored in a dry place and is ready for use.
2
Charging interval
Every 3 months
Battery with SN < 20000
every 5 months
Battery with SN ≥ 20000
every 9 months