Export configuration
The Export configuration function allows you to export all the
configuration settings made on the device to an SD card.
When exporting, all the configuration settings (including the
options) are transferred with the exception of the following
configuration settings:
Date and time
Serial number
Device runtime
Filter runtime
Date of last function check
Date of last maintenance
Number of start-ups
Import configuration
The Import configuration function allows you to import the
configuration settings exported to an SD card from one device
onto a second device.
Following the import, the second device is configured in exactly the
same way as the original device. The access code for the operator
menu is also adopted.
If you do not wish to adopt the customer-specific password for the
operator menu, you have two options:
Before the export: Reset the password to 0000 and export the
configuration.
Prior to import: Set the password to 0000 before exiting the
operator menu.
Configuration imports are saved in the log files. Configurations
can only be transferred between devices with the same software
version. Options subject to a charge are only imported if these
options are already activated.