Parameter
pMax
pMax CPR
CPR mode
Start mode
Metronome
Metronome freq.
etCO
/
(only
2
with capnography
option)
Possible values
10 mbar - 65 mbar
10 mbar - 65 mbar
Manual
100/min - 120/min
☑
☐
Description
Here you can set the maximum
ventilation pressure.
Here you can set the maximum
ventilation pressure in CPR mode.
Here you can set with which
submenu the CPR mode should be
started. If the CCSV option is
enabled, once the patient group
Child or Adult is selected, the device
always starts in continuous mode
with CPR CCSV.
Here you can activate or deactivate
the audio output of the metronome.
Here you can set the frequency of the
metronome tone.
Here you can determine whether an
alarm should be output in the event
of rising or dropping end-expiratory
CO
.
2
MEDUMAT Standard
6 Operator menu
2
EN
135