Weinmann MEDUMAT Standard2 Instructions For Use Manual page 132

Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

6 Operator menu
Parameter
Suppress maintenance
Allow filter counter reset
Dust load on filter
CO
unit (only with capnography
2
option)
Supply gas
Trigger setting
132
EN
MEDUMAT Standard
Possible values
Activated
Deactivated
Yes
No
100%
150%
200%
vol%
kPa
mmHG
93% O
2
100% O
2
3 levels
Units
2
Description
When maintenance becomes due, you can
suppress the maintenance notice within a
maintenance cycle one time for a maximum of
180 days from the date maintenance is due. The
device then behaves as if the maintenance were
not due in terms of the reminders. It is not
possible to reverse the suppression of the
maintenance notice within a maintenance cycle.
After the 180 days have elapsed, the
maintenance notice becomes active again until
the next service is performed and cannot be
deactivated again.
Even when the maintenance notice has been
suppressed, maintenance is still required
(see "11.2 Intervals", page
Here you can determine whether the user is
allowed to reset the hygiene filter at any time
during the function check.
Here you can set the load caused by
environmental factors (e.g., dust) for the
hygiene filter.
With an average load (100%) the hygiene filter
filter is able to function for approx. 24 hours of
ventilation in Air Mix mode, or for 6 months.
Here you can select which unit of measurement
the CO
values should be displayed in.
2
Here you can set the type of supply gas.
Here you can set the inspiration and expiration
trigger:
Simple three-level setting
Normal multi-level setting (in units)
224).