Loma IQ2 Service Manual page 32

Hide thumbs Also See for IQ2:
Table of Contents

Advertisement

Step Five
If you are using any of the serial links, then you will need to configure the serial
link. Configuration information can be found in the
section of this chapter.
page 28
Step Six
For detectors using IQ2v1.1 software and above go to the Loma menu and
follow the
Exit the Service menu by pressing the [
Step Seven
The machine will now be looking for metal in your product, although the
sensitivity will not be fully optimised, because the detector has not yet been
calibrated for the product. Before any products can be calibrated, it will be
necessary to perform a ferrite tune in order to minimise the effects of vibration
when running products with no product effect (see section on Principles of
Operation in your User Manual for more information).
The "TUNE FOR RESISTIVE" wand used to perform a ferrite tune is located
inside the control box. Perform a ferrite tune as follows:
Change to product F (the Access ferrite entry in the User Options menu must be
set to YES). To do this press the [
shown, now press the
To calibrate "Ferrite F" a Full calibration cycle must be completed. Before doing
this check to see if the conveyor is fitted with a photoeye prior to the detector, if
it is then return to the Configuration menu and set the Photo eye to NO (the
speed will require re-entering before you will be allowed to exit this menu). This
does not need to be done if there is no photoeye fitted. To perform a calibration,
press the [calibrate
shows "Full calibrate". Pressing the
When the "Pass product" message appears on the display, move the end of the
"Tune for Resistive Wand" with the ferrite piece back & forth through the centre
of the aperture at a rate as near as possible to the product speed until the
calibration cycle is completed. A temporary NON-METALLIC extension to the
wand will prevent hands, rings or watches entering the aperture.
Module 7026
ADC Setup menu
on page 22
" " " "
[
] key to accept the setting.
!
] key once, followed by the [
Configuration and Set up Information
Serial link configuration
.
C
] key four times.
!
] key, then the [
] key until "Ferrite F" is
!
] key twice, the display now
" " " "
[
] key initiates the calibration cycle.
on
5-10