Message
Bin full
Link corrupted
Link failure
PV test timeout
Note: Osgo = OScillator GOing.
PVS = Performance Validation System.
Module 6022
Cause
The reject bin full sensor has
detected that the bin is full.
Data corruption has occurred
between the detector and
LomaLink PC/other controlling
device. The operation of the
detector is unaffected.
A break in communication
between the detector and
LomaLink PC/other controlling
device. The operation of the
detector is unaffected.
A prompted for PVS test has
either not been done or
completed with the window time
set.
Fault and Error Diagnosis
Solution
If the bin is not full:
1. Check the Bin full sensor is
working correctly.
2. Check the Bin full sensor is
wired in the correct sense.
1. Check cabling and link
connectors from remote device.
2. Check LomaLink PC/ other
controlling device for reported
errors.
1. Check cabling and link
connectors from remote device.
2. Check connections between
the detector and LomaLink
PC/other controlling device for
loose or broken
wiring/connectors
1. Clear system fault.
2. Perform a PVS Test.
4-5