Loma IQ2 Service Manual page 29

Hide thumbs Also See for IQ2:
Table of Contents

Advertisement

Table of Tolerance settings available:
AUTOCAL:
Short
Long
Extra Long
Short is recommended for most applications, provided that at least eight packs
can be passed during a calibration cycle.
Standard, full size rectangular or square
5" to flow (and CF) freefall
129mm catalogue freefall
173mm catalogue freefall
219mm catalogue freefall
Handtmann style units
6000/7000 slimline combo
20mm pharmaceutical
38mm pharmaceutical
Ferrous in Foil
Table 1 Computation of coil gap by head type and minimum aperture dimension
Step Three
The detector must now be set up for the desired user options. Unless you are
completely familiar with the operation of the detector, it is recommended that all
the options be left at their default values.
Press the [
key enough times so that the display shows
explanation of these options is listed in the Definitions section of this chapter.
Press the
your options are desired:
Module 7026
Menu position
MIN
LOW
MED
HI
MAX
Select the desired auto-calibration period:
Photo Eye
8 packs
20 packs
N/A
Head Type
C
] key once to move out of the Configuration menu, and then the [
" " " "
[
] key once to move into the User Options menu, and then set up
Configuration and Set up Information
% of calibrated threshold added
15 % of calibrated threshold.
25 % of calibrated threshold.
50 % of calibrated threshold.
75 % of calibrated threshold.
100 % of calibrated threshold.
No Photo Eye
20 seconds
1 minute
2 minutes
Coil Gap
0.8 x minimum aperture dimension in mm
76mm
93mm
125mm
174mm
50mm
76mm
25mm
30mm
76mm
"User
options"
$
]
, an
on page 34
5-7