Loma IQ2 Service Manual page 27

Hide thumbs Also See for IQ2:
Table of Contents

Advertisement

Step One
If the unit is not already at the top of the Configuration menu, displaying the +Not
detecting+ warning message, go to this menu now by following this key
sequence:
1.
Press the [
2.
Press the [
3.
Press the
4.
Press the [
5.
Press the
6.
Press the down key three times, so that the display shows "Configuration".
7.
Press the [
8.
The display should now be showing "+Not detecting+".
Step Two
The correct parameters must now be entered into the configuration menu. To
gain access to the parameters use the [
parameter, then press the [
keys will alter the parameter (notice how the " " moves to the right and
becomes either "<" or ">""when the display is in parameter edit mode). Once the
correct parameter has been entered, simply press the [
the value.
Language: Select, the language required, from the following:
Type:
Terminology
Module 7026
#
] key once.
!
] key until the display shows "Service".
" " " "
[
] key once.
$
] key until the password shows "76".
" " " "
[
] key once. The display should now show "User options".
" " " "
] key once.
" " " "
] key to accept the command. The [
ENGLISH (default), FRANÇAIS, NEDERL, DEUTSCH, ESPAÑOL,
ITALIAN, PORTUGES, KATAKANA.
Select
CONVEYOR,
depending upon your system configuration.
FREEFALL:
Used when product is (drop through) gravity fed.
PIPELINE:
Used when product is pumped (in line) through a
piped system.
Configuration and Set up Information
$
!
] and [
] keys to move to the
" " " "
FREEFALL,
PIPELINE
$
!
] and [
]
] key again to save
or
PHARM
5-5